9.2.8.2 Loading configuration and calibration data into the device
NOTICE
All data which can be configured in the Setup menu are overwritten!
If the file is loaded into several devices, changing the network settings and the host
name is indispensable!
9.3 Resetting the device to the factory settings using the front-panel keys
Note:
Reset to the factory settings is possible only, when the CAL switch is open. The IP
address and the Hostname remain unaffected.
1. Open the CAL switch; see Chapter
7.1.3.1.
2. Click on [Select File] (depending on the Internet browser).
3. e.g.on the notebook, navigate to the folder where the backup file was saved.
4. Select Backup file.
5. Click on [Restore].
The selected file is loaded into the device.
6. Click on the symbol in the Internet browser to return to the WEB menu.
7. Close the CAL switch.
1. Press .
SEtuP appears on the display.
2. Press .
bIoS appears on the display.
3. Press OK.
FLASH follows on the display.
bIoS... appears on the display.
4. Press 2×.
dEL.EAr. appears on the display.
5. Press OK.
ErAS.E? and then no appear on the display.
6. Press .