7.5.2 Recalibrating the internal weighing point using the front-panel keys
Note:
If linearization is active (see Chapter 7.14.11), Error 91 appears on the display.
Example:
Max (maximum load) 2000.0 g
Scale interval 0.5 g
Set the dead load with empty scale.
Note:
The device has the following default setting when recalibration is started:
Max 3000 kg
Scale interval 1 kg
Start
Recalibrating (Cd 000)
Defining the max (maximum load) and weight unit (Cd 001)
1. Switch on the device; see Chapter 7.2. Allow for the warm-up phase for the device to
be completed.
2. Open the CAL switch; see Chapter 7.1.3.1.
3. Press
.
SEtuP appears on the display.
4. Press OK.
The "Cd" calibration menu flashes.
Cd 000 appears on the display.
5. Press OK to select the menu item.
CALIb appears on the display.
6. Press OK.
nEW (for new calibration) appears on the display.
7. Press OK.
Cd 001 appears on the display.