17.5.4 Sensitivity
Sensitivity 0.5 µV/e @ 10,000 e (class III); OIML R76
Max. resolution 7.5 million internal counts at 3 mV/V
Measurement input
(measuring signal + dead
load)
U
DC
= 0…max. 36 mV, symmetrical to zero
Dead load suppression max. measuring signal of U
DC
= 36 mV (dead load + ran-
ge); entry/calibration via software
17.5.5 Connecting cables
Length of the connecting cable between junction box and instrument
Cable type PR 6135, PR 6135A max. 500 m – length of the load cell cable
Length of the connecting cable between weighing platform and instrument
Cable type LiYCY max. 500 m
17.6 Mechanics
17.6.1 Design
Metal housing made of aluminum with steel cover and rear panel, protection class
IP65 (front).
17.6.2 Weights
Net weight 1.45 kg
Shipping weight 2.00 kg
17.7 Documentation on the CD included
The documents and manuals listed in the appendix (see Chapter
18.3) can be found on the
PR 5410 CD.