Note:
If a linearization was carried out (see Chapter 7.14.11), the following note appears on the
display after the [Dead load at] line is selected:
Cannot be changed here while linearization is active.
Changes cannot be made while linearization is switched on.
Only deleting of the linearization points deactivates the linearization mode!
7.14.11 Linearization
The measurement range for a straight can be optimized by setting the linearization
points.
Requirements:
Calibration of Max and dead load was done.
7.14.12 Calculating the test value
The calculation of the test value is called by pressing the [CalcTest] softkey.
The maximum load (Max) is displayed with the ID
without a weight unit. The value
determined during calibration by pressing the [CalcTest] softkey after starting the test is
displayed.
Depending on the settings under
- [Weighing point] - [Calib] - [Param] - [Test mode]
the following is displayed by calling the Test with the key later on:
- with [Absolute] the maximum load
- with [Relative] the deviation from the test value
7 Getting started X3 Process Indicator PR 5410
Minebea Intec EN-177

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea Intec X3 and is the answer not in the manual?

Minebea Intec X3 Specifications

General IconGeneral
BrandMinebea Intec
ModelX3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals