2.4.1 Installation
The device is suitable for installation in control panels or cabinets.
The device has to be installed in an EMC-compliant manner, see Chapter 4.3.
Component Protection class Installation
Housing front IP65
Housing rear panel IP30
Suitable for any operating en-
vironment.
To ensure proper cooling of the device, make sure air circulation around the device is not
blocked. Avoid exposing the instrument to excessive heat, e.g. from direct sunlight, and
vibrations. The ambient conditions in Chapter 17.4.1 must be taken into account at all
times.
With outdoor mounting, make sure that adequate weather protection is provided (for
temperatures, see Chapter 17.4.1).
2.4.2 Opening the device
WARNING
Working on the device while it is switched on may have life-threatening
consequences.
When removing covers or parts using tools, live parts or terminals may be exposed.
Please note that capacitors in the device may still be charged even after disconnecting
the device from all voltage sources.
Disconnecting the device from the power supply.
This device contains electrostatically sensitive components. For this reason, an
equipotential bonding conductor must be connected when working on the open device
(antistatic protection).
2.4.3 Supply voltage connection
The device does not have a power switch. It is in operation as soon as the power is
connected.
X3 Process Indicator PR 5410 2 Safety instructions
EN-12 Minebea Intec