EasyManua.ls Logo

Minebea Intec X3 - Effect of ambient conditions

Minebea Intec X3
362 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
17.4 Effect of ambient conditions
17.4.1 Ambient conditions
Reference temperature 23 °C
Ambient temperature for operation -10…+50 °C
Ambient temperature "verifiable" -10…+40 °C
Power-on temperature 0…+50 °C
Temperature range
Limits for storage/transport -20…+70 °C
Moisture <95%, non-condensing (acc. to
IEC 60068-2)
Protection class Housing: IP30
Front panel: IP65
Height <2000 m
17.4.2 Electromagnetic Compatibility (EMC)
All data in compliance with NAMUR NE 21, EN 45501 and EN 61326.
High frequency electromagnetic fields
(80…1000 MHz)
EN 61000-4-3 10 V/mHousing
Electrostatic discharge (ESD) EN 61000-4-2 6/8 kV
Fast transients (burst) EN 61000-4-4 1 kV
Peak voltages (surge) 1.2/50 µs EN 61000-4-5 1/2 kV
Signal and control
lines
Conducted disturbances by high fre-
quency coupling (0.15…80 MHz)
EN 61000-4-6 10 V
Fast transient disturbances (Burst) EN 61000-4-4 2 kV
Peak voltages (surge) 1.2/50 µs EN 61000-4-5 1/2 V
Conducted disturbances by high fre-
quency coupling (0.15…80 MHz)
EN 61000-4-6 10 V
Voltage dips EN 61000-4-11 0/40%
20/200ms
Mains inputs
Mains failure link EN 61000-4-11 20 ms
17.4.3 RF interference suppression
Interference emission
pursuant to EN 61326, Limit class A, for indust-
rial areas
17 Technical data X3 Process Indicator PR 5410
Minebea Intec EN-355

Table of Contents

Related product manuals