Display Error and Possible Cause
Err 61 CAL switch is closed.
Err 62 Calibration cannot be started.
Err 63 The weighing point parameters could not be saved.
Err 64 The default SPAN could not be set.
Err 66 Weighing point is in use.
Err 67 Weighing point is not tared.
Err 68 No mechanical stability of the scale.
Err 69 Weighing point is not in use.
Err 70 Tare is active.
Err 71 A weight error occurs during calibration via the front panel if the
connection to the weighing point is interrupted.
Err 72 This message is displayed for internal weighing point
- during the warm-up phase.
- when the device is in the automatic taring mode.
- if the device has been switched on with the scale loaded.
Switch the device off and on again.
Err 73 Dead load <-0.1 mV/V
Err 74 Arithmetic overflow
Err 75 Current weight < dead load
Err 76 Value for "d" is too high.
Err 77 Value for "d" is too low.
Err 78 Max < calibration weight.
Err 79 The weight cannot be displayed with the desired unit.
Err 80 Dead load + Max is too high.
Err 81 Max and scale interval do not match (1, 2, 5 10, 20, or 50).
Err 82 Not sufficient µV/d for W&M (legal-for-trade).
Err 83 Digital filter: Measurement time >160 ms.
Err 84 Filter frequency: Measurement time <40 ms/F
cut
.
Err 85 Filter frequency F
cut
is >40.00 Hz
Err 86 xBPI weighing point is selected.
This weighing point can only be calibrated via VNC.
Err 87 Sequence no. was not found in alibi memory.
Err 88 Sequence no. cannot be changed because the alibi memory is activa-
ted.
Err 89 No EW-Com interface assigned.
16 Error messages X3 Process Indicator PR 5410
Minebea Intec EN-351