SPM address Data type R/W Function
B5
X40
X41
X42
X43
X44
BYTE
BOOL
BOOL
BOOL
BOOL
BOOL
R
R
R
R
R
R
ADC status
Measuring signal negative (error 7)
Measuring signal >36 mV (error 3)
Internal arithmetic error; CAL data are perhaps faulty (error 1)
No or too low sense voltage (error 6)
No communication with xBPI scale (error 9)
B6
X48
X49
X50
BYTE
BOOL
BOOL
BOOL
R
R
R
R
Command status
Command error
Command active
Network failure signal
B7
X56
X57
X58
X59
X60
BYTE
BOOL
BOOL
BOOL
BOOL
BOOL
R
R
R
R
R
R
Active status
Test mode active
Calibration active
Instrument is tared
Pendeo/Connexx only: parameter [Unbal. check deviat.]
Pendeo/Connexx only: operation with a simulated load cell
X64 BOOL R/W Standard only: Read/write marker bit 1
X65 BOOL R/W Standard only: Read/write marker bit 2
X66 BOOL R/W Standard only: Read/write marker bit 3
X72 BOOL R/W Switch D11 to net weight.
X112 BOOL W Zero device.
X113 BOOL W Tare device
X114 BOOL W Reset the tare of the device
X115 BOOL W Start the test mode
X116 BOOL W Finish the test mode
X117 BOOL W Reset the power fail signal
X118 BOOL W Set fixed tare weight D31 as tare
X119 BOOL W Store the current gross weight in the preset tare memory (D31)
X120 BOOL W EasyFill only: Start printout.
Standard only: Weight report
X121 BOOL W Reset error B19 = 0.
X123 BOOL W Read out current time and date.
X124 BOOL W Save (set) current time and date.
B16 SINT R Exponent
Number of decimal places
Example: 1.23 is displayed
Exponent: 2
B17 SINT R Weight unit 1 = mg, 2 = g, 3 = kg, 4 = t, 5 = lb, 9 = oz
13 SPM X3 Process Indicator PR 5410