Note:
During calibration, the multiple range function is always switched off.
Example:
Range mode: "Multiple range"
Range 1: 0…1000 kg (when calibrating set scale interval: 1 kg)
Range 2: 0…2000 kg (next highest scale interval: 2 kg)
Range 3: 0…3000 kg (next highest scale interval: 5 kg)
7.14.15.3 Multi-interval scale (Class III or single range scale Class I and II with variable scale interval)
The multi-interval scale is a scale with a weighing range that is divided into intervals. Each
interval range has a different scale interval, where the weighing range is automatically
switched depending on the load on the scale and also when the load is placed on/
removed from the scale.
When [Range mode] = [Multi-interval], the scale has up to three ranges with different
resolution.
1. Choose "Range mode" from the -[Weighing point]- [Calib]- [Param] menu.
2. Select "Multiple range" and confirm.
3. Setting Switch point from range 1 to 2: enter "1000 kg" for the Range limit 1.
4. Setting Switch point from range 2 to 3: enter "2000 kg" for the Range limit 2.
5. Press the
softkey to exit and save calibration.
X3 Process Indicator PR 5410 7 Getting started