The weight display header includes the current interval range (R1, R2, or R3), Max, Min,
and d (or e with instruments used in legal metrology) (Example: multi-interval scale in
range 2):
The parameters [Range limit 1] and [Range limit 2] are the interval ranges.
As soon as the displayed weight exceeds range 1, the next highest range with the next
highest scale interval becomes valid (1->2->5->10->20->50).
Note:
During calibration, the multi-interval function is always switched off.
Example:
Range mode: "Multi-interval"
Interval range 1: 0…1500 kg (when calibrating set scale interval: 1 kg)
Interval range 2: 1500…2900 kg (next highest scale interval: 2 kg)
1. Choose "Range mode" from the -[Weighing point]- [Calib]- [Param] menu.
2. Select "Multi-interval" and confirm.
3. Setting interval range 1: Enter "1500 kg" for range limit 1.
4. Setting interval range 2: Enter "2900 kg" for range limit 2.
5. Press the
softkey to exit and save calibration.
7 Getting started X3 Process Indicator PR 5410
Minebea Intec EN-185