The weight unit for the CAL weight may differ from the unit in the device. Conversion
is automatic.
Note:
Error messages, calibration see Chapter
16.4.
1. Press the [by load] softkey.
2. Place the CAL weight on the scale.
3. Enter the weight of the CAL weight.
4. Confirm the entries.
5. Press
to select the weight unit.
6. Confirm entries with or .
Weight value, weight unit and measuring signal in mV/V corresponding to this
value are displayed in the line [Calibrated at].
The verification is displayed by "Setting Span by load…".
X3 Process Indicator PR 5410 7 Getting started