EasyManuals Logo

Minebea Intec X3 User Manual

Minebea Intec X3
362 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #163 background imageLoading...
Page #163 background image
Number of
vessel feet
Load cells Inteco®/..: Enter number.
Local gravity Enter local gravity (standard: Hamburg
9.81379 m/s
2
)
Max Enter maximum load:
0.000010…<3000>…9999998 <kg>, t, lb, g,
mg, oz
Scale interval Enter scale interval (1 d): <1>, 2, 5, 10, 20, 50
displayed according to the decimal places at
Max and the weight unit.
Dead load Dead load: Weight of the empty scale
CAL weight Enter calibration weight: 0.000010…9999998
<kg>, t, lb, g, mg, oz
Corner
correction
Platform 1, platform 2 (only appears when the
number of load cells = 8).
OK, when carried out.
Modify Only used for minor changes (e.g. changing the
dead load). Otherwise, always use [New]!
Param Parameter settings
Measuretime Measurement time: The value cannot be
changed. It is specified by the Connexx module.
Digital filter Selection of the digital filter (filter
characteristic): <off> (no filter), Bessel, aperiod.
(aperiodic), Butterw. (Butterworth), Tcheby.
(Tschebyscheff)
Fcut Only if filter is not "off", enter the corner
frequency: 0.1–80.0 Hz
W&M Select legal-for-trade mode: <none>, OIML,
NSC, NTEP
Unbal.
Check deviat.
Unbalanced check deviation: the plausibility
check is activated when the average deviation
is >0%.
Entry: 0…100 %
Standstill time Enter the standstill time: 0.01 s…<0.50 s>…2.0 s
(range depends on the measurement time).
Standstill
range
Enter the standstill range:
0.00 d…<1.00 d>…10.00 d (range depends on
the measurement time).
Tare timeout Enter the timeout when there is no standstill:
0.1 s…<2.5 s>…25 s
Zeroset range ± Range of zero point when there is no
standstill. Input:
0.00 d…<50.00 d>…10000.00 d
Zerotrack
indic. range
Enter the zerotrack indication range:
0.00 d…<0.25 d>…10000.00 d
Zerotrack step Enter the zerotrack steps:
0.00 d…<0.25 d>…10.00 d
Zerotrack time Enter the zerotrack time interval: <0.0 s>…25 s
Overload Enter the weighing range above the maximum
capacity (Max), without error message:
0…<9 d>…999999 d
Min
7 Getting started X3 Process Indicator PR 5410
Minebea Intec EN-161

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea Intec X3 and is the answer not in the manual?

Minebea Intec X3 Specifications

General IconGeneral
BrandMinebea Intec
ModelX3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals