- In the menu [Weighing point A] - [xBPI-Scale] - [Setup] at [Configuration] -
[Weighing parameters] - [Confirming adjust.] was set to "manual".
- The communication between the device and platform is active.
Procedure:
1. Select -[Weighing point]- [xBPI-Scale] and confirm.
2. Press the [Setup] softkey.
The parameters of the xBPI-scale are read into the device.
3. Select [Calibration]- [Adjust with user weight] and confirm.
An input window appears. The previously stored user weight is displayed.
4. Change the weight value if necessary using the keyboard and confirm.
The calibration process is carried out without a weight. The calibration status is
displayed.
5. Place the weight on the scale.
The deviation is displayed in the last line with increased resolution (10-fold).
6. Press the [Accept] softkey.
The weight is displayed in high-resolution (10x).
The data are saved and the instrument returns the following message:
7. Remove the weight.
8. Press the
softkey to exit the menu and to save the settings.
X3 Process Indicator PR 5410 7 Getting started
EN-196 Minebea Intec