Configuring digital outputs
3. Press multiple times to select "dl 174."
4. Press OK.
InP 1 appears on the display.
5. Press OK.
0001 appears on the display.
6. Press the cursor keys to enter the desired SPM address (see Chapter 13.4).
7. Press OK.
dI 175 appears on the display.
8. Press OK.
InP 2 appears on the display.
9. Press OK.
0001 appears on the display.
10. Press the cursor keys to enter the desired SPM address (see Chapter 13.4).
11. Press OK.
dI 176 appears on the display.
12. Press OK.
InP 3 appears on the display.
13. Press OK.
0001 appears on the display.
14. Press the cursor keys to enter the desired SPM address (see Chapter 13.4).
15. Press "Exit" to exit the Setup menu.
16. Press
.
SEtuP appears on the display.
17. Press OK.
The "Cd" calibration menu flashes.
Cd 000 appears on the display.
18. Press
multiple times to select "do 170."
19. Press OK.
8 Production X3 Process Indicator PR 5410