Each entry contains:
Sequence number 1–999999
Date YYYY/MM/DD
Time hh:mm:ss
Scale address Always A
Weight type G = gross, N = net, T = tare, PT = preset tare
Weight with unit
Depending on the configuration, the gross and/or net and/or tare weight are saved
(using the front-panel keys:
oP 063 see Chapter 7.5.1).
To view the data records using the VNC program, see Chapter 7.19.4.
To view the data records using the front-panel keys, see Chapter 7.6.1.
The alibi memory can be completely deleted, as long as the CAL switch is open, using the
front-panel keys only (see Chapter 7.6.2).
Example:
Data record as CSV file
1. Click on the [Retrieve alibi memory] menu item in the WEB menu.
The CSV file appears as an icon at the bottom of the screen.
2. Click on the icon.
Commas are used in the CSV file to separate the individual elements of an entry.
The file automatically opens, e.g., in Microsoft Excel.
3. Carry out a conversion in Microsoft Excel to see only one element per column. The
sequence number is placed in front of a consecutive line number.
9 Extended functions X3 Process Indicator PR 5410
Minebea Intec EN-299

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea Intec X3 and is the answer not in the manual?

Minebea Intec X3 Specifications

General IconGeneral
BrandMinebea Intec
ModelX3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals