ExpressionMR400InstructionsforUseGettingStarted3‐29
Simulation Mode
The MR400 is equipped with a simulation mode that displays computer generated data for
training or demonstration. As a safety feature, Simulation is displayed and appears on all
printouts while in simulation mode. Do not attach a patient to the MR400 when in simulation
mode and never activate simulation mode when a patient is connected. The MR400 will not
monitor patients while in the simulation mode. Activating simulation mode when a patient is
connected will result in a lapse in patient monitoring and could result in a delay in
treatment.
AllowstheMR400tooperateusinginternally‐
generateddatainsteadofpatientdata.
ToopentheSimulationModemenu
PresstheSetupkeyandthentheMonitorkey.On
theMonitorSetupmenu,selectService(Bio‐Med).
OntheService(Bio‐Med)menu,locateSimulation
ModeandselectOn.Enterthepassword.
The
followingcontrolsareavailable:
• Offdisplaysnormalmonitoringfunctions.
(Default)
• Ondisplayssimulationsofmonitoring
functions.
Toentersimulationmode
Toexitsimulationmode,turnoffthepowerswitch.
Step Action
1EnsurethatnopatientisconnectedtotheMR400.
2PresstheSetupkeyandthentheMonitorkey.
TheMonitor
Setupmenu appears.Currentsettingsaredisplayed.
3OntheMonitorSetupmenu,selectService(Bio‐Med).
TheService(Bio‐Med)menuappears.Currentsettingsare
displayed.
4LocateSimulationModeandselectOn.
5WhenEnterPasswordisdisplayed,enterthecorrectsix‐digitcode.
Simulationwillbedi splayedinthesystemmessageareaof
the
simulatedscreen.