Water Trap Replacement
Thewatertrapisintendedtoprotectthe
AGENTsystemfromhumidity,secretions,
bacterialcontaminationanddust.Replacement
ofthewatertrap isnecessarywhenthe
contentsinthewatertrapreachitsfullline.
Always discard the water trap when it becomes filled. Do not attempt to clean or reuse the water
trap. Accidental water ingress into the system can affect the gas measurements.
• The water trap must be checked every 17 hours of use and replaced as necessary. (Dispose
of the trap according to your facility’s biohazard procedure.)
• For optimum fit and compatibility, use only Invivo (Royal Philips) specified parts.
Toreplacethewatertrap
Step Action
1
Ifasamplinglineisconnected,thenremovethesamplinglinefromthe
sampleportonthewatertrap.
2Whilesimultaneouslypressingbothreleasetabsonthewatertrap,
removethewatertrapfromthewatertrapreceptaclethendisposeof
thewatertrap
properly.
Release tabs
Water trap receptacle