424AlarmsExpressionMR400InstructionsforUse
Alarm Limit Factory Defaults
Intheeventofpowerloss,anyalarmlimitsettingsthatwerechangedwillbelost.Allsettingsthat
mayhavebeenmodifiedtosuitaparticularpatientshouldbeconfirmedbeforemonitoring.
Atpowerup,theMR400willautomaticallysetallalarmlimitsasdeterminedbythedefault
selectedin
theEditUserSettingsmenu;seepage315.Thefactorydefaultalarmlimitsarelisted
inthetablebelow.
Note
You are restricted from making changes to the factory default settings.
SEV (Et) Expired Sevoflurane Vol. % Off, 0.1 8.0 0.1 8.0, Off 1
SEV (Fi) Inspired Sevoflurane Vol. % Off, 0.1 8.0 0.1 8.0, Off 1
N2O (Fi) Inspired Nitrous
Oxide
Percent No low alarm 0 80 1
O2 (Fi) Inspired Oxygen Percent 18 100 18 100 1
Breath Phase and Gas Unit
Low Alarm Limit High Alarm Limit
Low and
High Limit
Separation
Minimum Maximum Minimum Maximum
Vital Sign or Parameter Unit
Adult Pediatric Neo
Low
Limit
High
Limit
Low
Limit
High
Limit
Low
Limit
High
Limit
Heart Rate BPM 45 160 75 160 90 210
Heart Rate - Extreme
Bradycardia
BPM202020202020
Heart Rate - Extreme
Tachycardia
BPM202020202020
SPO2 Percent 85 Off 90 Off 90 Off
SPO2 - Desat Percent 80 80 80 80 80 80
CO2 (Et) mmHg
kPa
15
2.0
60
8.0
15
2.0
60
8.0
30
4.0
45
6.0
CO2 (Fi) mmHg
kPa
No low
alarm
4
0.5
No low
alarm
4
0.5
No low
alarm
4
0.5
CO2 (Resp) RPM 4 40 4 40 30 70
P1 / P2 Systolic mmHg
kPa
65
8.7
190
25.3
70
9.3
120
16.0
70
9.3
100
13.3
P1 / P2 Mean mmHg
kPa
55
7.3
135
18.0
50
6.7
90
12.0
40
5.3
90
12.0
P1 / P2 Diastolic mmHg
kPa
40
5.3
125
16.7
40
5.3
70
9.3
35
4.7
50
6.7