ExpressionMR400InstructionsforUse SpecificationsA23
Offset ±1percent
O
2
InterferingGasEffects:
N
2
O
CO
2
Halothane
Enflurane
Isoflurane
Desflurane
Sevoflurane
Acetone
Ethanol
Helium
Methane
NitricOxide
<0.3vol%@80vol%N
2
O
<0.3vol%@5vol%CO
2
<0.3vol%@5vol%HAL
<0.3vol%@5vol%ENF
<0.3vol%@5vol%ISO
<0.3vol%@18vol%DES
<0.3vol%@8vol%SEV
<0.3vol%@1vol%Acetone
<0.3vol%@0.1vol%Ethanol
<0.3vol%@80
vol%HE
<0.3vol%@0.1vol%Methane
<0.3vol%@50ppmNO
OxygenSen sor,OperatingTemperature 15to35°C(59to95°F)
OxygenSen sor,ExpectedOperatingLife Productlabeledwithausebydate;15monthsfrommanufacturing
date(2500hoursat100percentO
2
).Exchangerecommendedevery
12months.
CO2
Resolution 1percent
Range 0to100percent
Bellows Respiration
RespirationRateMeasurementRange 0to60breathsperminute
RespirationRateResolution 1breathperminute
RespirationRateAccuracy ±1breathperminute
Temperature (Optional)
(Allmeasurementsmadewithorwithoutasterile jacket)
Channel One
Units Celsius(°C)orFahrenheit(°F)
Range 20.0°Cto44.0°C(68.0°Fto111.2°F)
Resolution 0.1°C(0.1°F)