6‐2MonitoringSPO2ExpressionMR400InstructionsforUse
Philips has verified the compatibility of the monitor, probe, and cable specified in the
Accessory List. The user should verify that only Philips accessories specified in the Accessory
List are used. Otherwise, patient injury can result.
If dropped, the wSpO2
module must be verified for correct operation before use; see page 14-12.
Refer to your facility's biohazard procedure for disposal of SPO2 attachments and probes when
they become unusable. Usually probes are disposed of as medical waste per facility procedures.
Patient Preparation for SpO2 Monitoring
WhenmonitoringSpO2,theSpO2attachment,thesiteselectedonthepatient,theSpO2
attachment’spositiononthepatient,andtheambientenvironmentwillimpacttheperformance
andoperationoftheparameter.
Selecting the Site and SpO2 Attachment
Whenapplyingtheclipsorgripstothepatient,sitepreparationandthepressureandalignment
oftheSpO2attachmentareimportantfactorstoconsider.Selectthemostappropriatelimbthat
bestfitstheattachment’ssize.Formeasurementstobeaccurateandreliable,theoptimumfitis
reachedwhenthefi ber
headwindowsontheattachmentopposeeachotherwhilecoveringskin
ornail.RefertotheinstructionsprovidedwiththeSpO2attachmentwhenselectingand
connectingthecliporgri p.