Philips Expression MR400 - Suspend Mode; Simulation Mode

Philips Expression MR400
418 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
224SystemOverviewExpressionMR400InstructionsforUse
Toenternormalmode
PresstheMainScreenkey.
AnyopenmenuwillcloseandanyhighlightedVSboxwillbedeselected(thatis,thenormalscreen
willbedisplayed).Seepage214foranillustration.
Suspend Mode
WARNING
Suspend mode should never be used to silence alarms or when a patient is being actively
monitored as a delay in treatment and possible patient injury could result.
Suspendmodesupportspatientclinicianinteractionwithoutnuisancealarms,whichisuseful
whereminimaluserinteractionisrequired(forexample,whileapatientisnotbeingmonitored,
duringtransitionswhenremovingthemonitorfromonepatientandconnectingitonanother,or
ifcertainadjustmentsarebeingmadeto
thedeviceorotherequipment).
Insuspendmode,currentpatientinformationisprovided,butwiththefollowingoperational
exceptions:
•Audiblealarmsaredisabled;
•ActiveautomaticNIBPmeasurementsaresuspended;
•DefaultinflationpressuresareusedforallmanualNIBPreadings;and,
•IfequippedwithanIP5andprinter,automaticprintoutswillnotbe
generated.
Toentersuspendmode
PresstheSuspendkey.
Suspendedwillbedisplayedatthecenterofthescre en.See page220foranillustration.
Toexitsuspendmode
PresstheSuspendkey.
Simulation Mode
Simulationmode,apasswordprotectedfunction,supportstrainingandtestingneedsby
displayinginternallygenerateddataforvitalsignwaveforms,numericvaluesandstatuses.In
simulationmode,allpatientmonitoringisdiscontinued.Seepage329formenudetails.