8‐26MonitoringInvasiveBloodPres sure ExpressionMR400InstructionsforUse
HR Source
Selectsthesourcethatproducestheheartrate,asdisplayedintheECGandSPO2VSboxes
(identicaltoandinteractivewithsameoptionintheECGandSPO2menus).
Thefollowingoptionsareavailable:
• Autosetsthesourceautomaticallyaccordingtothehighestpriority activeinputthatisfirst
toreportvalidpatientdata.Thepriorityranking(highesttolowest)isECG,P1,P2,SPO2
(providedthattheP1andP2channelshavebeenlabeledABP;seeSetLabelonpage8‐24
fordetails).Thesourcewillbecomeunavailablewhenithasproducednovaliddatafora
period
often(10)ormoreseconds.Thesystemexaminesthehighestpriorityactiveinput.
Ifnotfound,thesecond‐highestpriorityinputischosen,etcetera.Ifnonearepresent,
thenNoneisdisplayedastheheartratemeasurementnumeric.
• ECGsetsECGasthesource.(Default)
•ABPsetsABP
asthesource(ifnopressurechannelislabeledABP,awarningboxwillallow
automaticrenamingandselectionbeforeproceeding;alsoseeSetLabelonpage8‐24).
• SPO2setsSPO2asthesource.
Tosettheheartratesource
Grids
ControlsthepressuregriddisplayforIBPwaveforms.
Step Action
1 SelecttheP1(orP2)VSbox.
TheP1menu(ortheP2menu)appears.Currentsettingsare
displayed.
2 SelectHRSource.
TheHRSourcemenuappears.Thecurrentsettingishighlighted.
3 Selectthedesiredsettingfortheheartratesource:
Auto
ECG
ABP
SPO2
Thesource
ischanged.