EasyManua.ls Logo

Philips Expression MR400 - HR Source; Grids

Philips Expression MR400
418 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
826MonitoringInvasiveBloodPres sure ExpressionMR400InstructionsforUse
HR Source
Selectsthesourcethatproducestheheartrate,asdisplayedintheECGandSPO2VSboxes
(identicaltoandinteractivewithsameoptionintheECGandSPO2menus).
Thefollowingoptionsareavailable:
Autosetsthesourceautomaticallyaccordingtothehighestpriority activeinputthatisfirst
toreportvalidpatientdata.Thepriorityranking(highesttolowest)isECG,P1,P2,SPO2
(providedthattheP1andP2channelshavebeenlabeledABP;seeSetLabelonpage824
fordetails).Thesourcewillbecomeunavailablewhenithasproducednovaliddatafora
period
often(10)ormoreseconds.Thesystemexaminesthehighestpriorityactiveinput.
Ifnotfound,thesecondhighestpriorityinputischosen,etcetera.Ifnonearepresent,
thenNoneisdisplayedastheheartratemeasurementnumeric.
ECGsetsECGasthesource.(Default)
•ABPsetsABP
asthesource(ifnopressurechannelislabeledABP,awarningboxwillallow
automaticrenamingandselectionbeforeproceeding;alsoseeSetLabelonpage824).
SPO2setsSPO2asthesource.
Tosettheheartratesource
Grids
ControlsthepressuregriddisplayforIBPwaveforms.
Step Action
1 SelecttheP1(orP2)VSbox.
TheP1menu(ortheP2menu)appears.Currentsettingsare
displayed.
2 SelectHRSource.
TheHRSourcemenuappears.Thecurrentsettingishighlighted.
3 Selectthedesiredsettingfortheheartratesource:
Auto
ECG
ABP
SPO2
Thesource
ischanged.

Table of Contents

Related product manuals