6‐8MonitoringSPO2ExpressionMR400InstructionsforUse
SPO2 Waveform and VS Box
TheSpO2measurementsaredisplayedasawaveformintheVStraceareaofthescreenandas
numericinformationintheSPO2VSbox.Otherdata,includingSpO2‐relatedalarminformation,
arealsoprovidedinthisar eaofthescreen,asdetailedbelow.
8 SelectthePatientType.
SeeSelectingthe
PatientTypeonpage3‐11.
9Checkforanydi splayedSPO2
messages
Ifamessageispresent,followtherecommendedactiontoachieve
betterresults(seepage4‐26).
10 Placethemoduleascloseaspossibletotheboreopening.(Ifthe
modulecanbeplacedoutsidethebore,positioningattheiso‐
centerisnotnecessary.)
11 KeeptheSpO2probeand
moduleoutsidetheMRsystembore.
12 Placethemoduleonacushionedsurface.
13 CoverthepatientSpO2attachmentsitewithopaquematerial.
14 Duringmeasurement,checkthepatienttoensurethatthe
applicationsite hasapulsatileflowandthatthesitehasnot
changedinthickness(forexample,duetoedema)causing
an
improperfit.
Step Action