Philips Expression MR400 - Invasive Blood Pressure (Optional)

Philips Expression MR400
418 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ExpressionMR400InstructionsforUse SpecificationsA13
*Forkilopascals(kPa),allow±1leastsignificantdigittoaccommodateroundofferrorforcalculatedvalues.
A.
(FromISO80601255.FDArecommendedforahealthy40yearoldmale.)
B.
Alarm Limits
EndtidalCO2
Lower
Upper
Off,or5to60mmHg(Off,or0.7to8.0kPa)
5to90mmHg,oroff(0.7to12.0kPa,oroff)
FractionalinspiredCO2
Lower
Upper
Nolowalarm;
0to20mmHg,oroff(0to2.7kPa,oroff)
Respiration
Lower
Upper
Off,or4to40
RPM
20to100RPM,oroff
GAS or Vapor Halothane
Enflurane Isoflurane Desflurane Sevoflurane N
2
O
MAC Level, % vol fraction 0.77 1.70 1.15 6.00 2.10 105
Measured
Quantitative Effects of Gas or Vapor
Gas
N
2
O HAL ENF ISO SEVO Xenon Helium DES Ethanol Isopropanol Acetone Methane
Carbon
Dioxide
NE @
60%
NE
@
4%
NE
@
5%
NE
@ 5%
NE
@
5%
ME1
@
80%
NE @
50%
ME2
@
15%
NE @
0.1%
NE @
0.1%
NE @
0.1%
NE @
1%
No Effect (NE)
Minimal Effect 1 (ME1) = Negatively bias Carbon Dioxide values by up to an additional 5 mmHg at 38 mmHg
Minimal Effect 2 (ME2) = Concentrations greater than 5% will positively bias Carbon Dioxide values by up to an additional 3 mmHg at 38
mmHg
***Metered dose inhaler propellants: Unspecified
Invasive Blood Pressure (Optional)
Pressure Amplifier
IsolationVoltage 5KVDC
SignalRange 30to+250mmHg
Sensitivity 5µV/V/mmHg
GainAccuracy ±0.5percent
Bandwidth 0to10Hz(3dB)
CO2 (Optional LoFlo)