A‐14Specifications ExpressionMR400InstructionsforUse
TransducerOffsetRange ±300mmHg
Transducer (REF 989803179721)
OperatingPressure ‐50to300mmHg
OverpressureLimits ‐400to5000mmHg
Sensitivity 5µV/V/mmHg±1@6VDCand22°C(71.6°F)
ZeroOffset <25mmHg
ZeroDrift <2mmHgin8hours
InputImpedance 300to350ohms
Auto Zero
Range +300mmHg
ZeroAccuracy ±1.0mmHg
ResponseTime 1second,notifiesoperatorwhendone
Pressure Wave Display
NumberofChannels 0,1or2
ABP,PAPandLAP Numericdisplayofsystolic,meananddiastolicpressures
CVPandICP Numericdisplayofmeanpressureonly
Pressure Scale Ranges
(UserSelectable)
0to250mmHg
0to200mmHg
0to150mmHg
0to100mmHg
0to75mmHg
0to45mmHg
Pulse Rate (when derived from P1 or P2)
Range 30to250BPM
Accuracy ±2percentoffullscale
Resolution 1BPM
Alarm Delay
Transducer Disconnect Sixseconds
PressureDisconnect Sixseconds
HighandLowPressure Tenseconds
Alarm Limits
When“HR”isderivedfromP1orP2
Lower
Upper
Off,or30to250BPM
60to250BPM,oroff
Systolic,Mean,Diastolic
Lower
Upper
Off,or‐30mmHgto250mmHg(Off,or‐4.0to33.3kPa)
‐30mmHgto250mmHg,oroff(‐4.0to33.3kPa,oroff)
Invasive Blood Pressure (Optional)