ExpressionMR400InstructionsforUse MonitoringTemperature117
Tomakebodytemperaturemeasurements
Post-Measurement Processing
Aftermonitoringtemperature,processthesensorasfollows.
Step Action
1Ensurethatajackethasbeenplacedonthesensor(seePlacingthe
SensorinaJacket,onpage115).
2Ifneeded, applylubricanttothejacketforinsertionintothe
patient.
CAUTION CAUTION
Never use petroleum-based lubricants. A water-based lubricant (for
example, Surgical Lubricant, REF 989803168891) can be used to
facilitate insertion.
3Insertthesensingtipofthesensorintothepatientatan
appropriatedepth.
WARNING
Never insert the sensor beyond the patient segment of
the sensor. Insertion beyond the patient segment can
lead to difficulties removing the jacket from the
patient.
4Ifdesired,changetheunitofmeasure(Celsiusisthedefault
setting);seeTEMPMenu,onpage1110,fordetails.
5Performthemonitoringprocedure,allowingatleast2minutesfor
themeasurementtostabilize;seeTEMPVSBox,onpage118,for
details.
6 Aftertheprocedure,removethesensor
fromthepatient.
WARNING
Ensure that the entire jacket is removed from the
patient when withdrawing the sensor. Failure to do
so can potentially lead to jacket material being left
inside the patient.
7 Immediatelycleananddisinfectthesensor(seePostMeasurement
Processingonpage117).