A24Specifications ExpressionMR400InstructionsforUse
Accuracy ±0.5°C(±0.9°F)
Confirmingchangesinameasurementagainstothervitalsignmea
surementsshouldbestandardroutineduringuse.
ResponseTime Themeasuringtimetoobtainasteadystatereadingwithinthe
manufacturer’saccuracyspecification s iswithin15seconds,
complianttoISO80601256,Particularrequirementsforbasic
safetyandessentialperformanceofclinicalthermometersforbody
temperaturemeasurement.
NumericDisplayUpdateTime 2seconds
SensorType Fiberoptic,multipleuse(whenusedwithsingleusesterilizedjack
ets)
ApplicationSite Axillary,esophageal,rectal
MeasurementMode Direct
Alarm Limits
Lower
Upper
Off,or20.0to44.0°C(Off,or68.0to111.2°F)
20.0to44.0°C,oroff(68.0to111.2°F,oroff )
Non-Invasive Blood Pressure
Oscillometrictechnology(withaninflatablecuff)determinessystolic,diastolicandmeanarterialpressures
PatientTypes Adult,PediatricandNeonate
Pneumatic Systems
CuffInflationPressure Initial:165mmHg(22kPa)forAdult,130mmHg(17.3kPa)for
Pediatric,and100mmHg(13.3kPa)forNeonate;allpressuresare±
15mmHg(2kPa)
SubsequentinflationpressuresdeterminedbylastNIBP
measurement
OverpressureProtection Automaticcuffpressurereleaseifinflationpressureexceeds300
mmHg(40kPa)for
AdultandPediatricmodes,and150mmHg(20
kPa)forNeonatemode
UnitofMeasure Millimetersofmercury(mmHg)orkilopascals*(kPa)
Measurement Range
Syst olic
Adult
Pediatric
Neonate
30to270mmHg(4.0to36kPa)
30to180mmHg(4.0to24kPa)
30to130mmHg(4.0to17.3kPa)
Temperature (Optional)
(Allmeasurementsmadewithorwithoutasterilejacket)