ExpressionMR400InstructionsforUse SpecificationsA‐7
HeartRate Automaticmodeselectsthevitalsigntoprovidetheheartrate
accordingtovitalsignsourceavailabilityandpriority.Ifnosource
available(ifnovitalsignmeetsthecriteria,thentheheartrate
sourcewillbedisplayedasNoneandno
heartratewillbe
produced.
Manualmodeselectiontoprovidetheheartrateisalsoavailable.
PulseOximeter Pulserate,pulsewaveform(normalized)andpercentsaturation
Trends Heartrate,respirationrate,P1and/orP2(systolic,diastolic,
mean),NIBP(systolic,diastolic,mean),EtCO2,O2,N2O,SpO2,
andAgents
CO2 Bothend‐tidalandfractionalinspiredCO2
NIBP Pressures
(systolic,mean,diastolic)andstatus
RespirationRate Resp irationratederivedfrombellowsorCO2
N2O
EtN
2
OavailableinAgentMACbox
FiN
2
Onotdisplayed
O2 Inspired,end‐tidal(averagedpercent)
AGENT Automaticidentificationofprimaryandsecondaryagents
(Desflurane,Isoflurane,Enflurane,HalothaneorSevoflurane)
displayingbothend‐tidal(Et)andfractionalinspired(Fi)
concentrations.
Temperature Bodytemperature(°Cor°F)
ECG
ECG Amplifier
Protectedagainstdefibrillatorandelectrosurgerypotentials
StandardLeadConfigurations I,II,III,AVR,AVL,AVF
LeadFail Passive,sensingsignalimbalance
ECGInputImpedance >2.5MΩ,single‐ended(accordingtoIEC60601‐2‐27,50.102.3)
ElectrodeContactImpedance 20Kohms@10Hz
Heart Rate
Resolution 1beatperminute(BPM)
PulseRateRange 30to250BPM(Adult)
30to300BPM(Neonate,Pediatric)
Accuracy ±1percentor±1 BPM,whicheverisgreater.
Displayed Information