6‐18MonitoringSPO2ExpressionMR400InstructionsforUse
HR Tone Source
Setsthesourceusedfortheheartratetone(identicaltoandinteractivewithsameoptioninthe
intheMonitorSetup>SoundAdjustmenuandintheECGmenu).
Thefollowingoptionsareavailable:
• Offremovestheheartbeatdetectedsymbolfromthedisplayandnopulsetonewill
be
sounded.(Default)
• QRSprovidestheheartbeatdetectedsymbolandatonetriggeredbytheQRSdetection
fromtheECGvitalsign.
• SPO2providestheheartbeatdetectedsymbolandatonemodulatedbytheSPO2vital
sign,wherethelowertheSPO2value,thelowerthepitch.
Tocontroltheheart
ratetonesource
3 Selectthedesiredsettingfortheheartratesource:
Auto
ECG
ABP
SPO2
Thesourceischanged.
Step Action
1 SelecttheSPO2VSbox.
TheSPO2menuappears.Currentsettingsaredisplayed.
2 SelectHRToneSource.
TheHRToneSourcemenuappears.Thecurrentsettingis
highlighted.
3 Selectthedesiredsettingforthetone
source:
Off
QRS
SPO2
Thesettingischanged.
Step Action