614MonitoringSPO2ExpressionMR400InstructionsforUse
•10%
•20%
•40%
•60%
•80%
•100%(Default)
ToadjustthesizeoftheSPO2waveform
SeeChangingtheSPO2WaveformAmplitudeonpage610.
Averaging Time
Selectshowquicklythereadingrespondstochangesinthepatient’ssaturation,whereselecting
alongerdurationwillpreventthesaturationvaluefrom changingquicklywhichcanbeusefulfor
avoidingalarmtriggeringinpatientswithverydynamicconditionssuchasneonataland
pediatrics.
Thefollowingoptions(inseconds)areavailable:
•5Sec
•10Sec(Default)
•15Sec
TosettheaveragingtimeoftheSPO2reading
Step Action
1 SelecttheSPO2VSbox.
TheSPO2menuappears.Currentsettingsaredisplayed.
2 SelectAveragingTime.
TheAveragingTimemenuappears.Thecurrentsettingis
highlighted.
3 Selectthedesiredtimeforaveraging:
5Sec
10Sec
15Sec
The
settingischanged.