ExpressionMR400InstructionsforUse MonitoringInvasiveBloodPressure8‐9
catheterfilledwithI.V.solutionpriortoinsertion.Followcathetermanufacturer’sinsertion
instructions.
VI. Fast Flushing
A. Activatethefastflushmechanismofthecontinuousflushdeviceandcheckdripchamberto
confirmfastflush.
B. FOLLOWINGEACHFASTFLUSH,DRIPCHAMBERDROPRATEMUSTBEOBSERVEDTO
VERIFYCOMPLETECLOSURE.
VII. Checking for Leaks
Afterapproximately1minutehaselapsed,theflowrateshouldbeobservedatthedripchamber
toensurethatthecontinuousflushdeviceisoperatingproperly.Avisualinspectionforleaks
shouldalsobemadesince asmallleakcanmisrepresenttheactualcontinuousflowthroughthe
catheter.Aprotocolshould
beestablishedaccordingtothehospitalstandardofcareforroutinely
checkingthesystemforproperfluidsourcepressure,flowrateandleaks.
VIII. In the MR Room
• The IBP transducer must not be allowed past the 5,000 gauss line, or transducer
failure, inaccurate readings, noisy MRI images or patient burn may result.
• Never place the pressure transducer’s stopcocks or port covers within 8 cm (3.2
inches) of the field of view of the MR bore as inaccurate readings or noisy MRI images
can result.
• An offset occurs when the pressure transducer is repositioned in the magnetic field.
The transducer must be zeroed prior to the MRI examination after the transducer is in
its final setup position. Moving the transducer after zeroing may cause a measurement
offset to occur.
IntheMRroom,ensurethatthetransducerislevelwiththeheartthatthetransducerinterface
cablingisnotloopedortouchingthepatient(refertheWarningsonpages8‐3and8‐4),andthen
re‐zerothetransducer;seepage8‐16.
Non-physiological pulsatile P1 (or P2) waveform (for example, those found during intra-
aortic balloon pump use) can lead to inaccurate blood pressure readings. If questionable
values are observed, recheck the patient’s pressures by alternate means before administering
medication or therapy.