530MonitoringECGExpressionMR400InstructionsforUse
SPO2setsSPO2asthesource.
Tosettheheartratesource
HR Tone Source
Setsthesourceusedfortheheartratetone(identicaltoandinteractivewiththesameoptionin
theMonitorSetup>SoundAdjustmenuandintheSPO2menu).
Thefollowingoptionsareavailable:
Offremovestheheartbeatdetectedsymbolfromthedisplayandsoundsnopulsetone.
(Default)
QRSprovidestheheartbeatdetectedsymbolandatonetriggeredbytheQRSdetection
fromtheECGvitalsign.
SPO2providestheheartbeatdetectedsymbolandatonemodulatedbytheSPO2vital
sign,wherethelowertheSPO2value,thelowerthepitch.
Tocontroltheheartratetonesource
Step Action
1 SelecttheECGVSbox.
TheECGmenuappears.Currentsettingsaredisplayed.
2 SelectHRSource.
TheHRSourcemenuappears.Thecurrentsettingishighlighted.
3 Selectthedesiredsettingfortheheartratesource:
Auto
ECG
ABP
SPO2
Thesourceischanged.
Step Action
1 SelecttheECGVSbox.
TheECGmenuappears.Currentsettingsaredisplayed.