1210MonitoringNIBPExpressionMR400InstructionsforUse
Mean Format
1
2
3
4
10
5
6789
Item Name Definition
1 Alarm flag area Displays NIBP alarm flags when detected; see page 4-31.
2 Mean upper alarm limit
Is the upper limit setting for the mean alarm, and accesses the NIBP Alarm
Limits menu
3 Mean lower alarm limit Is the lower limit setting for the mean alarm, and accesses the NIBP Alarm
Limits menu
4 Mean numeric Is the detected patient’s mean pressure measurement
5 Diastolic numeric Is the patient’s diastolic pressure measurement
6 Systolic numeric Is the patient’s systolic pressure measurement
7 Unit of measure Indicates that the NIBP measurement numeric values are given in mmHg
(millimeters of mercury) or kPa (kilopascals); see Pressure Units in the
System Config menu on page 3-30.
8 Elapsed time / cuff pressure Is the time since the last completed NIBP reading, in the following format:
<hh>:<mm>:<ss>, where
<hh> = Two-digit hours field
<mm> = Two-digit minutes field
<ss> = Two-digit seconds field
Note
During a reading, this displays the cuff pressure.