EasyManuals Logo

Philips Expression MR400 User Manual

Philips Expression MR400
418 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #162 background imageLoading...
Page #162 background image
438AlarmsExpressionMR400InstructionsforUse
Other Status Indications
Message, Location,
Indication
Condition
All Alarms Are Off
System message area, no
alarm light, no alarm tone
All alarm limits have been turned off.
Audio Off
System message area, no
alarm light, no alarm tone
Alarm sound is silenced.
Audio Paused
System message area, no
alarm light, no alarm tone
Alarm sound is paused.
Change NIBP Cuff
System message area, no
alarm light, no alarm tone
The Patient Type was changed so the NIBP cuff should be changed.
Check Alarm Vol
System message area, no
alarm light, no alarm tone
Power was just turned on, settings were recalled.
status information pane, blue
alarm light, INOP alarm tone
The charge level is low for the wECG module battery (or batteries). Install at least one charged
battery into the wECG module.
WARNING
A red battery symbol indicates that the module batteries have fallen below
the required operational output and module shutdown with loss of
monitoring will occur. Immediately replace the module batteries to avoid a
loss in monitoring.
Status information pane, no
alarm light, no alarm tone
No communication between the MR400 and wECG module. Ensure that the wECG module is set
to the same network channel as the MR400 cart. If both are the same, use an alternate setting.
If the indication persists, contact technical support or authorized service personnel.
ECG Test Signal
System message area, no
alarm light, no alarm tone
The ECG Test Signal is on.
Status information pane, no
alarm light, no alarm tone
No communication between the MR400 and IP5. Ensure that the IP5 is set to the same network
channel as the MR400 cart. If both are the same, use an alternate setting.
If the indication persists, contact technical support or authorized service personnel.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips Expression MR400 and is the answer not in the manual?

Philips Expression MR400 Specifications

General IconGeneral
BrandPhilips
ModelExpression MR400
CategoryMedical Equipment
LanguageEnglish

Related product manuals