ExpressionMR400InstructionsforUseAlarms4‐39
Status information pane, blue
alarm light, INOP alarm tone
Charge level of the cart batteries is low. Connect the MR400 cart to external power and allow the
batteries to charge.
A red battery symbol indicates that the main batteries in the MR400 have fallen
below the required operational output and system shutdown with loss of monitoring
will occur. Immediately locate an AC outlet and connect the MR400 to avoid a loss in
monitoring.
Overscale
ECG alarm flag area, no
alarm light, no alarm tone
The size of the ECG waveform is too large and the tops of the ECG waveforms are clipped (that is,
cut off). Reduce the size using the Scale setting; see page 5-28.
Real Tones Disabled
System message area, no
alarm light, no alarm tone
Normal audio sounds are suspended as a volume adjustment is in progress.
Simulation
System message area, no
alarm light, no alarm tone
The system is in simulation mode. Turn off power to exit this mode.
Status information pane, blue
alarm light, INOP alarm tone
Charge level of the wSpO2 module battery is low. Install a charged battery into the wSpO2 module.
A red battery symbol indicates that the module batteries have fallen below
the required operational output and module shutdown with loss of
monitoring will occur. Immediately replace the module batteries to avoid a
loss in monitoring.
Status information pane, no
alarm light, no alarm tone
No communication between the MR400 and wSpO2 module. Ensure that the wSpO2 module is set
to the same network channel as the MR400 cart. If both are the same, use an alternate setting.
If the indication persists, contact technical support or authorized service personnel.