118ImportantInformationExpressionMR400InstructionsforUse
Attaching the SpO2 Probe to the wSpO2 Module
TheSpO2probe,necessaryfortakingSpO2relatedmeasurementsusingthewSpO2module,
mustbeconnectedprioruse.
WARNINGS
Only perform this attachment at a location outside of the MR magnet room. Failure to
observe this warning may result in serious injury.
Connecting an other than specified SPO2 probe to the wSpO2 module can cause
inaccurate SPO2 readings and damage the module.
ToattachtheSPO2probe
InserttheSPO2probeconnectorintotheDB9connectoronthewSpO2
modulethensecurelytightenbothscrews.
5LocatethebatteryswitchandtoggleitintotheOn(I)position.
6 Reinstalltheservicepanelcover,andsecureittotheWPUusing
thetwoscrews.
7Replacetheshieldcap.
Thiscompletestheinstallati on andconnectionofthemainand
reservebatteries.ConnectACmainspowerthenallowthese
batteriestochargeforatleast12hoursbeforeinitialuse;seepage
121.
Step Action
Battery switch
1 wSpO2 module
2 DB-9 connector
3 SPO2 probe connector
4 Screws
3
4
2
1