422AlarmsExpressionMR400InstructionsforUse
Offdoesnotdisplaythealarmlimitsettings.
Ondisplaysthealarmlimitsettings.(Default)
Tocontrolthedisplayfunctionforthealarmlimitsettings
Adjustable Alarm Limit Ranges
Withtheexceptionofbellowsr e spiration,eachparameteralarmhasadjustablelimitsasindica tedin
thetablesbelow
.
Notethatthealarmlimitnumericvaluescanbesettooff,withtheexceptionof
O2,N2O,FiCO2.
TheMR400alsopreventscrossoveroflowandhighalarmlimitsettings,anda
minimumnumberofunitsseparatestheselowandhighsettings.Whenaparameterhasbeen
turnedof f, itsalarmlimitswillbeoff.Alarmlimits areadjustablebythesameresolutionspecifiedin
eachparameter'smeasuremen tresolution
setforthinAppendixA.
Note
The minimum and maximum values for the low and high limits represent the most extreme settings
possible. For some vital signs, these values can be obtained for a low or high alarm limit only if
the other limit is off.
ThetablebelowprovidesthealarmlimitrangesfortheMR400
.
Step Action
1PresstheSetupkeyandthentheAlarmskey.
TheAlarmsmenuappears.Currentsettingsaredisplayed.
2 SelectLimitsDisplay.
TheLimitsDisplaymenuappears.Thecurrentsettingis
highlighted.
3 Selectthedesiredoptionfordisplayofthealarmlimitsettings:
Off
On
Thesettingisselected.
Vital Sign or
Parameter
Unit Patient Type
Low Alarm Limit* High Alarm Limit* Low and
High Limit
Separation
Minimum Maximum Minimum Maximum
HR BPM All Off, 30 250 60 250, Off 2
SPO2 Percent All Off, 50 100 70 100, Off 2
CO2 (Et) mmHg
kPa
All Off, 5
Off, 0.7
60
8.0
5
0.7
90, Off
12.0, Off
2