EasyManua.ls Logo

Philips Expression MR400 - CO2 Waveform and VS Box

Philips Expression MR400
418 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ExpressionMR400InstructionsforUse MonitoringCO2(LoFlo Option)77
CO2 Waveform and VS Box
TheCO
2
measurementisdisplayedasawaveformintheVStraceareaofthescreenandas
numericinformationintheCO2VSbox.Otherdata,includingCO2relatedalarminformation,are
7Placetheairwayadapterattheproximalendoftheairwaycircuit.
CAUTION
Always insert the patient sampling line into the CO2 port before
inserting the airway adapter into the breathing circuit. Failure to
follow this may introduce a leak in the circuit, thereby reducing set
minute volume.
Note
Do not place the airway adapter between the ET tube and the elbow
as this may allow patient secretions to accumulate in the adapter. If
pooling does occur, replace the airway adapter. To prevent moisture
from draining into the airway adapter, always place the adapter
tubing in a up position, as shown above.
8 SelectthePatientType.
SeeSelectingthePatientTypeonpage311.
9Checkthattheconnectionshavebeenmadecorrectlybyverifying
thepatient’sbreathingeffortswiththedisplayedwaveform.
WARNING
Before completion of patient setup, ensure that the
patient’s breathing efforts coincide with the displayed
CO2 waveform.
10 Ifthefollowingwarningisdisplayed:
PersistentCO2occlusiondetected.
PleaseclearocclusionandpressOK
toresumeusingCO2
Clearany pinchesorobstructionsinthesamplinglinebefore
proceeding.
Step Action

Table of Contents

Related product manuals