1410MaintenanceandTroubleshooting ExpressionMR400InstructionsforUse
Sterilization
TheMR400cart,wirelessmodulesandaccessoriesarenotsterilizable;donotimmerseanypartof
theseitemsinfluidorattempttocleanthemwithunspecifiedliquidcleaningagents.Severe
damage,notcoveredbythewarranty,willresult.
5Cleanthecartandmodulesbythoroughlywipingthedevicesusing
CaviWipes
disinfectanttowelettes.Discardtheusedtowelettes
(refertoyourfacility'sbiohazardprocedurefordisposal).
Note
Follow the Instructions for Use from the disinfectant manufacturer
to clean the cart and wireless modules.
6Disinfectthecartandmodulesbythoroughlywettingthedevices
usingCaviWipesdisinfectanttowelettes.Discardtheused
towelettes(refertoyourfacility's biohazardprocedurefor
disposal).
Note
Follow the Instructions for Use from the disinfectant manufacturer
to disinfect the cart and wireless modules.
7
Allowthecartandthewirelessmodulestodry.(Norinsingis
required.)
8Checkthecartandthewirelessmodulesforanyresidualdebris.If
anydebrisispresent,repeatsteps3 through6thenreexaminethe
cartandwirelessmodulesbeforeproceeding.
9Checkthecartandwirelessmodules
fordamaged,looseormissing
hardware.Contacttechnicalsupportifnecessary.
10 Storethemodule;seepage212fordetails.
Step Action