ExpressionMR400InstructionsforUseImportantInformation123
Note
The reserve batteries cannot be removed, but can be switched off. For instructions on the complete
removal of power to the MR400, see page 14-2.
Wireless Module Batteries
ModulebatteriesprovidepowertothewECGandwSpO2modules.Modulebatteriesare
interchangeable,nonmagnetic,andcanbehandledsafelyintheMRmagnetroom.
CAUTION
To minimize the chance of image artifacts, never place module batteries in the MRI field of view.
2Pressthebatteryejectbuttontopartiallyejectamainbatteryfrom
thebatterycompartment,andthengraspthebatteryandpullto
removeitcompletelyfromtheMR400.
(Ifthebatterydoesnotrelease,applyaslightforwardpressureto
thebatterywhi le pressingthebatteryejectbutton.)
3
Repeatsteps1and2toremovetheothermaincartbatteryonthe
oppositesideoftheMR400.