4‐4AlarmsExpressionMR400InstructionsforUse
Audible Alarm Indications
Do not rely exclusively on the audible alarm system for patient monitoring. Adjustment of
alarm volume to a low level or off during patient monitoring may result in patient danger.
Remember that the most reliable method of patient monitoring combines close personal
surveillance with correct operation of monitoring equipment.
Dependingupontheseverityandtypeofthealarmcondition,alarmprioritytoneswithdistinct
pitchesaresoundedbythespeaker(seeillustrationonpage4‐3).Whenmultiplealarmsare
detected,thehighestpriorityalarmisannounced.
Onceactivated,anaudiblealarmindicationcontinuesaslongasthealarm
conditionispresentor
untilyouacknowledgethealarmbypressingtheAudioPausekeyortheAlarmkey(orbyplacing
theMR400insuspendmode).
Alarm Sound State Indication
Dependinguponthecurrentsetting,oneofthreepossiblesymbolsareprovidedtoindicatethe
alarmsoundstate,whichisdisplayedatalltimesintheinformationbar(seepage2‐14).Symbol
definitionsforthealarmsoundstateareprovidedinthetablebelow.
Alarm priority Fixed Pitch Mnemonic Notes
High c c c – c c
Medium c c c
INOP e c
Displayed Symbol Alarm Sound State
Alarm audio armed
Alarm audio paused
Alarm audio off