ExpressionMR400InstructionsforUse SpecificationsA‐25
Mean
Adult
Pediatric
Neonate
20to255mmHg(2.7to34kPa)
20to160mmHg(2.7to21.3kPa)
20to120mmHg(2.7to16kPa)
Diastolic
Adult
Pediatric
Neonate
10to245mmHg(1.3to32.7kPa)
10to150mmHg(1.3to20kPa)
10to
100mmHg(1.3to13.3kPa)
Accuracy
PressureMeasurementAccuracy Maximum meanerror±5mmHg(±0.6kPa)withastandarddeviation
oflessthan8mmHg(1kPa)
PressureMeasurementResolution 1mmHg(0.1kPa)
PressureTransducerRange 0to300mmHg(0to40kPa)
Modes
Manual Immediateuponoperatorcommand
Automatic Determinationsautomaticallymadewithselectableintervalsof1,2,
3,5,10,15,20,and30minutes
Non-Invasive Blood Pressure