A22Specifications ExpressionMR400InstructionsforUse
FiIsoflurane
Lower
Upper
Off,0.1to5.0
Vol. %
0.1to5.0Vol. %,Off
EtEnflurane
Lower
Upper
Off,0.1to5.0
Vol. %
0.1to5.0Vol. %,Off
FiEnflurane
Lower
Upper
Off,0.1to5.0
Vol. %
0.1to5.0Vol. %,Off
EtSevoflurane
Lower
Upper
Off,0.1to8.0
Vol. %
0.1to8.0Vol. %,Off
FiSevo flu rane
Lower
Upper
Off,0.1to8.0
Vol. %
0.1to8.0Vol. %,Off
EtDesflurane
Lower
Upper
Off,0.1to18.0
Vol. %
0.1to18.0Vol. %,Off
FiDesflurane
Lower
Upper
Off,0.1to18.0
Vol. %
0.1to18.0Vol. %,Off
FiO
2
Lowerandupper 18to100percent
O2
Resolution 1percent
Range 0to100percent
SignalOutput(atconstanttemperatureand
pressure)
10mV±1.5mV@20°C/20.95%O
2
MaximumResponseTime(21%to100%step
changethroughpatientsamplinglineasseen
inWPUgasmonitorwindow)
Adult/Pediatric<7.3seconds
Neonate:<8.2seconds
Accuracy(includesstabilityanddrift),full
scale*
±1%at0to40%
±2%at40to60%
±3%at60to80%
±4%at80to100%
*Gas
measurementperformancerequirementsaremetafterthemaximumwarmupperiod.
AGENT (Optional)