430AlarmsExpressionMR400InstructionsforUse
Violated secondary
agent Et value
AGENT The end-tidal measurement for the
secondary agent gas has violated
an alarm limit setting.
Flashing yellow secondary agent Et numeric in
the AGENT VS box, yellow alarm light,
medium priority alarm tone, where OVR
displayed if the value is greater than the highest
specified
Violated secondary
agent Fi value
AGENT The fractional inspired
measurement for the secondary
agent gas has violated an alarm
limit setting.
Flashing yellow secondary agent Fi numeric in
the AGENT VS box, yellow alarm light,
medium priority alarm tone, where OVR
displayed if the value is greater than the highest
specified
Missing primary and
secondary agent data
AGENT The primary and secondary agent
data, once present, can no longer
be produced.
Flashing yellow dashes (- - -) in the primary and
secondary numerics in the AGENT VS box,
yellow alarm light, medium priority alarm tone
Violated N2O value AGENT The N2O measurement has
violated an alarm limit setting.
Flashing yellow N2O numeric in the GAS VS
box, yellow alarm light, medium priority alarm
tone
Missing N2O data AGENT The N2O data, once present, can
no longer be produced.
Flashing yellow dashes (- - -) in the N2O
numeric in the GAS VS box, yellow alarm light,
medium priority alarm tone
Violated O2 value AGENT The O2 measurement has violated
an alarm limit setting.
Flashing yellow O2 numeric in the GAS VS
box, yellow alarm light, medium priority alarm
tone
Violated O2 value AGENT The O2 measurement is less than
18 percent.
Flashing red O2 numeric in the GAS VS box,
red alarm light, high priority alarm tone
Missing O2 data AGENT The O2 data, once present, can no
longer be produced.
Flashing yellow dashes (- - -) in the O2 numeric
in the GAS VS box, yellow alarm light, medium
priority alarm tone
Missing bellows
respiration rate data
RESP The bellows respiration data, once
present, can no longer be
produced.
Flashing yellow dashes (- - -) in the respiration
rate numeric in the RESP VS box, yellow alarm
light, medium priority alarm tone
Violated temperature
value
TEMP The temperature measurement
has violated an alarm limit setting.
Flashing yellow temperature numeric in the
TEMP VS box, yellow alarm light, medium
priority alarm tone, where OVR displayed if
the value is greater than the highest specified
or UND if the value is less than the lowest
specified
Missing temperature
data
TEMP The temperature data, once
present, can no longer be
produced.
Flashing yellow dashes (- - -) in the
temperature numeric in the TEMP VS box,
yellow alarm light, medium priority alarm tone,
where OVR displayed if the value is greater
than the highest specified or UND if the value
is less than the lowest specified
Violated non-invasive
blood pressure systolic
value
NIBP The non-invasive blood pressure
systolic measurement has violated
an alarm limit setting.
Flashing yellow systolic numeric in the NIBP VS
box, yellow alarm light, medium priority alarm
tone, where OVR displayed if the value is
greater than the highest specified or UND if
the value is less than the lowest specified
Patient Alarm From Condition Indication