ExpressionMR400InstructionsforUse MonitoringInvasiveBloodPressure821
Tochangetheinvasivebloodpressurealarmlimitsettings
Note
See chapter 4 for detailed alarm limit setting instructions and options.
7 Current adjustment
8P1 Systolic Alarm Limits menu label (active adjustment shown)
9 Alarm limit, minimum
10 Lower P1 (Sys) alarm limit setting
11 Upper P1 (Sys) alarm limit setting
12 Alarm limit, maximum
Step Action
1 SelectthealarmlimitsettingsintheP1(orP2)VSbox.
TheP1AlarmLimits(ortheP2AlarmLimits)menuappears.
Currentsettingsaredisplayed.
2 Selectthedesiredpressuresetting:
P1(Sys)
P1(Dia)
P1(Mean)
Thesettingisselected.
3 SelecttheLowbutton
ortheHighbutton.
Theselectedbuttonwillbehighlightedandthecurrentadjustment
willbedisplayed.
4Usingthekeypad,ortheincrement,decrement,orOffbuttons,
enterthedesiredsetting.
Thecurrentadjustmentwillreflectthechange.
5PresstheEnterbuttontosavethesetting.
Thealarmlimitsettingis
updated.
6
Tochangetheremainingsettings,repeatsteps2,3,4and5.
Thecurrentadjustmentwillreflectthechange.