ExpressionMR400InstructionsforUse MonitoringAgentsandGases(AGENTOption)97
TopreparetheAGENTtubing
Step Action
1Ensurethatthewatertraphasbeeninstalledinthewatertrap
receptacle(seepage913forinstallationinstructions).
2InserttheLuerlockconnectoronthesamplingline(REF94018)
intothesample
portonthewatertrapandthentightenthe
connector(nomorethanonehalfturnshouldberequired).
3 Accordingthetypeofpneumaticci rcuitrequiredtoperformAGENT
monitoring,connectthepneumaticcircuititems(seethediagrams
above).
WhereequippedwithaLuerlock,onlyahalfturnofthe
Luerlock
connectorshouldberequiredwhereequipped;otherwise,ensure
thatanytubingconnectorhasbeenpressedfirmlyontothe
associatedadapter.
4Verifythateachconnectionistightbyholdingthenasalprongs(or
thepatientairwayadapter)closetoyourearandlisteningfora
hissingsound.
Increasethe
flowtemporarilyifnecessaryandthenreducetothe
prescribedflowrateoncetheflowisverified.
5Onadailybasis,performasystemtest;seepage98fordetails.
6Applythesamplinglinetothepatient(seepage910).
Sample port
Sampling line