ExpressionMR400InstructionsforUse MonitoringAgentsandGases(AGENTOption)911
Toapplythesamplinglineairwayadapter
3Passthetubingovertheearsandbehindthehead,
ensuringthepatient’sheadwillnotrestonanypart
ofthecannulawhilethepatientislyingdown.
4Slidethesleeve
towardthepatient’sheadtoassurea
goodfitofthecannula.
5 SelectthePatientType.
SeeSelectingthePatientTypeonpage311.
6Checkthattheconnectionshavebeenmadecorrectlybyverifying
thepatient’sbreathingeffortswiththedisplayedCO2waveform.
WARNING
Before completion of patient setup, ensure that the
patient’s breathing efforts coincide with the displayed
CO2 waveform.
7 AfterallowingtheAGSsystemtorunforatleast1minute,pinchor
sealthesamplinglinefor5secondsandverifythatOcclusionis
displayed.
Ifthismessagedoesnotappear,checkalltubingconnectionsfor
leakageandretest.
Step Action
1Ensurethatthesamplinglineisclean,dryand
undamaged.Replace
thesamplinglineifnecessary.
Step Action