16ImportantInformationExpressionMR400InstructionsforUse
MECHANICALSAFETY
Warning
WARNING
Do not remove covers from this product unless expressly instructed to do so in this
Instructions for Use. Moving parts are present within this product. Removing covers could
lead to serious or fatal personal injury.
Covers should normally only be removed by qualified and authorized service personnel. In this
context, qualified means those legally permitted to work on this type of medical electrical product
in the jurisdiction(s) in which the product is being used, and authorized means those authorized by
the user of the product.
FIRESAFETY
Useofanelectricalproductinanenvironmentforwhichitwasnotdesignedcanleadtofireor
explosion.Fireregulationsforthetypeofmedicalareabeingusedshouldbefullyapplied,
observedandenforced.Fireextinguishersshouldbeavailableforbothelectricalandnon
electricalfires.
Warning
WARNING
Only use extinguishers on electrical or chemical fires, which are specifically labeled for those
purposes. Using water or other liquids on an electrical fire can lead to fatal or other serious
personal injury.
Ifitissafetodoso,attempttoisolatetheproductfromelectricalandothersuppliesbefore
attemptingtofightafire.Thiswillreducetheriskofelectricshocks.
ELECTROSTATICDISCHARGE
Electrostaticdischarge(ESD)canamounttoasignificantvoltage,whichmaycausedamageto
PCBsorothersystem
components.
Caution
CAUTIONS
Always wait at least ten seconds after the product is switched OFF before switching the
product back to ON.
Always use proper static procedures, protection, and product prior to opening and during
handling of this product. This product contains components that are electrostatic sensitive.
Failure to use ESD procedures may cause damage to these components. Such damage to
components is not covered by Philips warranties.
ConnectionstosensitivepartsareidentifiedbytheESDwarningsymbol(see
inset).ESDdamageiscumulativeandmaynotbeapparentatfirst,asindicated
byahardfailure,butcancausedegradedperformance.Therefore,alwaysuse
properESDhandlingprocedures.ESDcanresultfromlowhumidityconditions,
useof
electricalequipmentoncarpeting,linens,andclothing.