ExpressionMR400InstructionsforUseImportant Information1‐9
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The Expression MR400 MRI Patient Monitoring System is intended for use in the electromagnetic environment
specified below, and the customer or the user should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF Emissions, CISPR 11 Group 1 The Expression MR400 MRI Patient Monitoring System uses RF
energy only for its internal functions. Therefore, its RF emissions
are very low and are not likely to cause any interference in
nearby electronic equipment.
RF Emissions, CISPR 11 Class B
The Expression MR400 MRI Patient Monitoring System is
suitable for use in all establishments, other than domestic
establishments and those directly connected to the public low-
voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic Emissions, IEC
61000-3-2
Class B
Voltage Fluctuations / Flicker
Emissions, IEC 61000-3-3
Complies