EasyManua.ls Logo

Philips Expression MR400 - Page 316

Philips Expression MR400
418 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
124MonitoringNIBPExpressionMR400InstructionsforUse
Thebladderwidthofthecuffmustbe40
percentofthecircumferenceofthelimb.Itis
alsoadvisabletokeeptheairvolumetoa
minimumbyusingthesmallestcuffsize
possibleforeachpatient.Thepointof
maximumoscillationsiscoincidentwithmean
arterialpressureregardlessofarterialelasticity
solongastheratioofairvolumeinthecuff to
thevolumeofthearteryundercompression
doesnotgreatlyexceedten(10)toone(1).
ForacorrectNIBPcufffit:
•Adultandpediatricpatients—theindex
lineonthesizechosenshouldfallwithin
therangelinewhenplacedonthe
patient.
Neonatalpatients—thesizechosen
shouldbewithinthestated
circumferencerangeforthelimbofthe
neonate.
•Allpatients—alignthecufftoensurethe
arterymarkisplacedovertheartery.
•Allpatients—themiddleof
thecuff
shouldbeplacedattheleveloftheright
atriumoftheheart.
WARNING
Single use devices, as indicated on the device packaging, should be disposed of after use and
must never be reused. Follow your hospital’s guidelines for appropriate disposal. Reuse of
single-use devices can result in spread of patient infection, degradation of monitoring
performance, or inaccurate measurements.
Index line
Range line
Circumference
range
Artery mark
NIBP cuff

Table of Contents

Related product manuals